CONTEC MS100 MS200 MS400 Simulator
MS200 Blood Pressure Simulator It is a multi – purpose test device for testing NIBP Monitor adopted Oscillometry. Concise and elegant in appearance, easy and convenient to operate, and it can generate the pressure up to 400mmHg (53.3 kPa) by the built-in air pump. With the functions of dynamic BP simulation under several conditions, static pressure calibration, automatic leak test and over-pressure relief test, which can quickly and accurately analyze each performance of NIBP Monitor.
Features
Provides pressure gauge testing,pressure leak testing,relief valve testing functions
Build-in pump provides pressure source with leak and pressure relief tests
NIBP simulations for both arm and wrist cuffs.
Provides adult,neonate, eight variations patients conditions,user-defined simulations.
Provides arrhythmias,respiratory artifact simulations.
Built-in air chamber to simulate adult and neonate cuff.
65k colors,4.3″ LCD display.
Technical index
Pressure measurement:
Units:mmHg,kPa,cmH2O,inH2O,psi
Range:0mmHg-400mmHg
Resolution:1 mmHg,0.1 kPa,1 cmH2O,1 inH2O,0.1 psi
Accuracy:0-300 mmHg:±0.5% of reading ±1 mmHg;
301-400 mmHg:±2%of reading
Pressure Generator:
Static Pressure Range:50 mmHg-400 mmHg
Resolution:1mmHg
Difference between target pressure and actual pressure :±10 mmHg from 100-400 mmHg with a minimum volume of 300 cc
Internal leak rate:
<2 mmHg per minute, with a minimum volume of 300 cc
Pressure Leak Test:
Pressure generate up to 400 mmHg (53.3 kPa) with internal pump. The screen always show the peak pressures and the real-time pressures. Leak rate displayed after 10 seconds.
Pressure Relief Test:
Pressure generate up to 400 mmHg (53.3 kPa) with internal pump, Displays peak pressure.
Standard BP Simulations:
Preset blood pressure:
1#: BP: 120/80(93) HR: 80 Pulse Volume: 0.68
2#: BP: 150/100(116) HR: 80 Pulse Volume: 0.65
3#: BP: 200/150(166) HR: 80 Pulse Volume: 0.60
4#: BP: 255/195(215) HR: 80 Pulse Volume: 0.55
5#: BP: 60/30(40) HR: 80 Pulse Volume: 0.75
6#: BP: 80/50(60) HR: 80 Pulse Volume: 0.71
7#: BP: 100/65(76) HR: 80 Pulse Volume: 0.69
Patient Conditions:
BP Simulations
Healthy: BP: 120/80(93) HR:75 Pulse Volume: 0.68
Weak Pulse: BP: 110/80(90) HR:95 Pulse Volume: 0.50
Mild Exercise: BP: 140/90(106) HR:120 Pulse Volume: 1.00
Strenuous Exer: BP: 140/90(106) HR:162 Pulse Volume: 1.40
Obese: BP: 120/80(93) HR:90 Pulse Volume: 0.50
Geriatric: BP: 150/110(123) HR:95 Pulse Volume: 0.40
Tachycardia: BP: 120/105(110) HR:130 Pulse Volume: 0.40
Bradycardia: BP: 120/60(80) HR:45 Pulse Volume: 1.10
Arrhythmia Simulations:
Premature Atrial Contract 1
Premature Atrial Contract 2
Premature Ventricular Contract
Atrial Fib and PVCs
Respiratory Artifacts
Spontaneous Breathing 1
Spontaneous Breathing 2
Spontaneous Breathing 3
Controlled Ventilation
Neonate simulations
User can select External or Internal Neonate;
1#: BP: 35/15 HR: 120bpm Pulse Volume: 0.11
2#: BP: 60/30 HR: 120bpm Pulse Volume: 0.10
3#: BP: 80/50 HR: 120bpm Pulse Volume: 0.10
4#: BP: 100/70 HR: 120bpm Pulse Volume: 0.10
Wrist simulations
1#: BP:120/80(93) HR: 80bpm Pulse Volume: 0.50
2#: BP:160/100(120) HR: 80bpm Pulse Volume: 0.50
3#: BP:80/55(63) HR: 80bpm Pulse Volume: 0.50
User-Defined simulations
Allows the operator to define blood pressure simulations.
Systolic(mmHg) 20-250
Diastolic(mmHg) 10-200
Heart Rate,(bpm) 30-250
Pulse Volume,(cc) 0.1-2.4
Simulant characteristics:
The maximum pulse volume:2.4cc
The maximum heart rate:250bpm,200bpm at pulse volume>2.00cc.
Internal Adult cuff volume:20cc
Internal neonate cuff volume:300cc
Blood pressure units: mmHg,kPa
Heart Rate Accuracy:±1BPM
Patient Conditions:
Weak pulse, Tachycardia,Obese,Geriatric ±1%±1BPM
Mild exercise±1.5%±1BPM
Strenuous exercise±3%±1BPM
Display
65k colors 4.3 inch TFT Display, brightness can be changed.
Power
AC100 – 240 V 50/60 Hz, 60 VA
Environmental Conditions:
Working temperature: 15℃ to 40℃(59℉ to 104℉)
Storage temperature: -20℃ to +65℃(-4℉ to +149℉)
Relative humidity: < 90%,No condensation.
Size (L x W x H):27 cm x 26.5 cm x 13.2 cm
Weight: About 4kg
Package included:
1XMS200 NIBP Simulator
1XPower cord
2XFuse
1XUser ‘s Manual